A Chinese walker supplier's FDA registration is verified from two separate FDA records, not from a certificate on a sales deck. Confirm the establishment registration and the device listing in FDA's public database, match both to the legal entity on your invoice, then check the annual renewal date (21 CFR Part 807).
What Is the Difference Between FDA Registration and Device Listing?
Two separate records exist, and a supplier can hold one without the other (21 CFR Part 807). Establishment registration records the facility and the activities performed there. Device listing records each device placed in commercial distribution, with its product code and listing owner. Both live in the same public database, and both are queried by the establishment name rather than by a certificate number.
Registration is annual. FDA opens the renewal window from 1 October to 31 December each year, and an establishment that misses the window shows as expired even if the facility has not changed. Listing is per device and is updated when a device or its product code changes. Neither record is a finding of conformity, and neither is issued after a technical review of the product.
| Record | What it records | Held by | Does not prove |
|---|---|---|---|
| Establishment registration | Facility, owner or operator, activities performed | The establishment that manufactures or processes the device | That the device is cleared or legally marketed |
| Device listing | Each device, its product code, listing owner | Establishment or specification developer | That the facility passed an inspection |
| US agent designation | US contact for a foreign establishment | Foreign establishment | Product conformity or performance |
Source: 21 CFR Part 807, FDA establishment registration and device listing database
A third record explains most surprises. For devices made to a US brand owner's specification, the listing may sit with the specification developer rather than the factory you are auditing. The factory can then be completely legitimate and still not appear as the listing owner. This is normal contract manufacturing, and it is the reason a name search alone is not a verification.
How Do You Run the Five-Step Verification?
The five steps run in order, and each produces a document you keep in the order file. Step one is to fix the legal entity name and address exactly as they appear on the quotation, then identify the product code the device is listed under. Without the exact registered name, the database search returns a trading company, a group holding company or nothing at all.
Step two searches the establishment record and captures the registration status, the activities performed and the expiry. Step three matches the device listing to that establishment and confirms the product code and the models you intend to buy. Step four reconciles the entity across registration, listing, invoice, label and US agent designation. Step five confirms renewal currency and writes down what the record does not cover.
| Step | Action | Evidence to keep | Pass condition |
|---|---|---|---|
| 1 | Fix legal entity name and product code | Quotation, business licence, label artwork | Name matches the registered name exactly |
| 2 | Search the establishment record | Database result with status and expiry date | Status active at the order date |
| 3 | Match the device listing | Listing record showing product code and models | Purchased models fall inside the listing |
| 4 | Reconcile entity across documents | Invoice, label, US agent designation | One legal entity across all documents |
| 5 | Confirm renewal and scope | Renewal date, model-to-listing mapping | Renewal current, scope covers the order |
Source: 21 CFR Part 807, in-house order verification procedure
The output of the five steps is not a yes or no answer. It is a short file containing five dated screenshots or extracts, one sentence stating what the record covers, and a list of what remains unverified. That file is what protects a buyer during a customs query or a customer audit, months after the sales conversation has ended.
Which Inconsistencies Stop a Verification?
Most failed verifications are consistency failures rather than missing records. A registration that exists, a listing that exists and an invoice that names a third company is a failed verification even though every individual record is genuine. The mismatch is the risk, because it means the entity that sold you the goods is not the entity the record describes.
The pattern is common with trading companies that source from several factories. Their own registration may cover warehousing and distribution, while the manufacturing record sits with each factory. That structure is workable, but it must be documented: which entity holds the manufacturing record, which entity issued the declaration, and which entity appears on the label.
| Inconsistency | What it usually means | Action |
|---|---|---|
| Registration under a different legal entity than the invoice | Trading company sells, factory or brand owner holds the record | Obtain a written explanation and authorisation to reference the record |
| Listing without the product code you are buying | Device sold under a different classification or not listed | Hold the order until the listing is confirmed |
| Registration expired at the order date | Annual renewal missed | Do not accept until renewal is shown |
| Purchased models outside the listing | Listing covers a subset of the catalogue | Request a model-to-listing mapping |
| No US agent on a foreign establishment | Import documentation gap | Request agent details before booking freight |
Source: 21 CFR Part 807, in-house order verification procedure
One inconsistency deserves separate handling: a supplier that presents a registration number as a product approval. Registration numbers identify establishments, not products, and they carry no statement about performance, testing or market authorisation. A supplier who cannot explain that distinction is unlikely to produce a usable technical file on request.
What Should Be Written Into the Purchase Order?
Write the registration evidence into the purchase order, because it is the only point where you can make it a condition of payment. The clause should name the documents, the entity that must issue them and the moment they are due. A clause that asks for compliance documents in general terms is difficult to enforce when the shipment is already at the port.
Three requirements cover most of the exposure. First, the establishment registration record and the device listing, dated within the current renewal cycle. Second, a statement of which legal entity is the legal manufacturer and which holds each record. Third, the documentation set supplied per order: declaration, test reports and labelling artwork for the purchased models.
| Clause element | What to specify | Why it belongs in the order |
|---|---|---|
| Registration evidence | Establishment record and device listing, dated | Creates a payment condition rather than a promise |
| Entity declaration | Legal manufacturer, record holder, label entity | Prevents a three-entity mismatch at the border |
| Product code | Code the purchased models are listed under | Confirms the listing covers the order |
| Testing documentation | Reports per model, with revision and date | Supports the technical file request |
| Change notification | Written notice before a record, model or label change | Keeps the verification valid across repeat orders |
Source: 21 CFR Part 807 and 21 CFR 890, in-house sourcing terms
The registration and listing are statements of fact held by the manufacturing partners and provided per product and order scope. They are not a statement that a device meets a performance standard, and they do not replace test reports or a technical file review. Treatment of the two is kept separate in our quality and compliance documentation, and the same separation should appear in a buyer's supplier file.
FAQ
Do walkers carry FDA approval?
Walkers are generally Class I physical medicine devices, and many are exempt from premarket notification. FDA does not issue a marketing approval for that device group; it registers establishments and lists devices. A supplier claiming product-level approval for a standard walker is describing something the framework does not provide.
Can a Chinese factory be FDA registered without a 510(k)?
Yes. Premarket notification and establishment registration are separate obligations. A Class I device can be exempt from premarket notification under 21 CFR 890 and still require establishment registration and device listing before it enters commercial distribution in the United States.
How often must FDA registration be renewed?
Annually, between 1 October and 31 December. The registration record shows a status and an expiry date, and a lapsed renewal appears as expired even when nothing at the facility changed. Buyers should re-check the record at each order rather than relying on a screenshot from a previous year.
Does a trading company's registration cover the factory?
Usually not. A trading company's registration covers its own activities, which may be warehousing or distribution. The manufacturing record sits with the factory or with the specification developer that owns the design. Ask which entity holds each record, and get that split in writing.
What is a US agent, and is one required?
A foreign establishment that registers with FDA must designate a US agent to act as its local contact (21 CFR 807.40). The agent receives correspondence and is available for FDA communication. An establishment record without an agent designation is an incomplete record for a foreign factory.
What this guide does not cover
- This guide is a verification method, not a legal opinion on any supplier, product or shipment. Whether a specific device may be imported rests with the buyer, the importer of record and their regulatory counsel.
- Customs entry, tariffs, labelling review by other agencies and state-level requirements are outside scope.
- 510(k) content, FDA inspection outcomes and warning letters are referenced only where they affect how a record should be read; they are not assessed for any supplier here.
Send the exact legal entity name and the models you are evaluating, and we will return the registration status, the listing scope and the documentation available per model within one business day. You can request a quote with your quantities, review the standard walkers range, or read how our quality and compliance documentation is released against a named model.
