Walkers and rollators reach the EU market as Class I medical devices under Regulation (EU) 2017/745, and most require no notified body involvement. What an importer must hold is a complete technical file, a signed EU Declaration of Conformity, CE marking on device and packaging, and EUDAMED registration (EU MDR 2017/745).
How Are Walkers and Rollators Classified Under EU MDR?
Most walkers and rollators fall under Class I because they are non-invasive and have no measuring function (EU MDR 2017/745, Annex VIII, Rule 1). Classification is decided by intended purpose, duration of use and whether the device measures something, not by price or material. Two features move a walking aid out of plain Class I: a measuring function, and sterile delivery. A rollator with an integrated scale is assessed as Class Im and needs a notified body for that function.
A Class I device without a measuring function is assessed without notified body involvement. The manufacturer compiles the technical documentation, signs the Declaration of Conformity and applies the CE marking. No third party reviews the file before the product ships. That is the commercial risk for an importer: the only gate between the factory and the EU border is the manufacturer's own paperwork.
| Device configuration | Rule applied | Resulting class | Notified body involved |
|---|---|---|---|
| Non-invasive, no measuring function (typical walker, rollator, cane) | Annex VIII, Rule 1 | Class I | No |
| Non-invasive with a measuring function (for example a built-in scale) | Rule 1 with Section 3.3 | Class Im | Yes, for the measuring function |
| Supplied sterile (rare for walking aids) | Rule 1 with Section 3.3 | Class Is | Yes, for sterility |
Source: Regulation (EU) 2017/745, Annex VIII, Rule 1 and Section 3.3
If a supplier describes a walker as Class I but the model includes any measurement output, the classification claim and the assessment route no longer match. Ask which rule was applied and read the intended-purpose statement before accepting the class stated in a quotation.
What Four Document Families Must an Importer Hold?
An importer should hold four document families before the first shipment: technical file, EU Declaration of Conformity, CE marking artwork, and EUDAMED registration. Each answers a different question, and a missing family cannot be covered by an email assurance from the sales contact. The four families are separate records, not four pages of one certificate.
The technical file is the evidence base: design and manufacturing information, general safety and performance requirements, risk management, verification and validation, and the clinical evaluation. The Declaration of Conformity is the legal statement that the device meets the regulation. CE marking is the visible output of that statement. EUDAMED registration makes the device and the economic operators traceable.
| Document family | What it establishes | Who issues it | Where the importer checks it |
|---|---|---|---|
| Technical documentation | Design, materials, test evidence, risk management, clinical evaluation | Manufacturer | File index reviewed against Annex II checklist |
| EU Declaration of Conformity | Legal statement of conformity and device identification | Manufacturer, signed and dated | Content review, UDI-DI and model list match |
| CE marking and labelling | The declaration is visible on device and packaging | Manufacturer | Artwork, label proof, IFU language set |
| EUDAMED registration and UDI | Device and economic operator traceability | Manufacturer, authorised representative, importer | EUDAMED device record and UDI-DI |
Source: Regulation (EU) 2017/745, Annexes II, III and IV, Articles 10 and 27 to 31
Two details decide whether this set is usable. First, the Declaration of Conformity must identify the device precisely enough to cover the models you are buying, including the UDI-DI. A declaration that names a product family without a model list is difficult to defend during a market surveillance check. Second, the CE marking must appear on the device or its sterile packaging and in the instructions for use, in the language of the member state of sale.
How Do You Verify a Declaration of Conformity Before Accepting It?
A Class I declaration is verified by content and consistency, because most walkers have no notified body certificate to check. Verification means comparing the declaration against the label, the invoice, the technical file index and the EUDAMED record. A single inconsistency is enough reason to hold the shipment until it is explained in writing.
Article 13 of the regulation makes this an importer obligation rather than good practice. The importer must verify that the device carries the CE marking, that the Declaration of Conformity has been drawn up, and that the manufacturer and, where applicable, the authorised representative are identified. Where a notified body is involved, the importer must also verify the certificate and its scope.
| Item to verify | What to compare it against | Common defect found |
|---|---|---|
| Legal manufacturer name and address | Commercial invoice and packaging label | Trading company named as the legal manufacturer |
| Device identification and UDI-DI | Label, packaging, EUDAMED record | Model list narrower than the models shipped |
| Classification and rule applied | Classification reasoning in the technical file | Class IIa claimed without a notified body number |
| Notified body number, where applicable | NANDO register entry and scope | Number present but expired or out of scope |
| Signature, date and revision | Current technical file revision | Undated declaration or superseded revision |
Source: Regulation (EU) 2017/745, Annex IV and Article 13
Certificate numbers for the two FDA and EU documents are not published by us in advance. Where a number is required for a tender or a customs file, request it per model and per order, and treat the supplied number as unverified until it is checked against the issuing register. This is the same discipline applied in our quality and compliance documentation, where documents are released against a named model rather than a catalogue page.
Who Is Responsible for What Along the Supply Chain?
Obligations split by role, and an importer carries duties that cannot be delegated to the Chinese manufacturer (EU MDR 2017/745, Article 13). A non-EU manufacturer must appoint an authorised representative established in the Union, and that representative is the counterpart a market surveillance authority will contact.
Importer duties include checking that the device is CE marked and that the declaration exists, that the manufacturer and representative are identified, that storage and transport conditions are met, that a complaint register is kept, and that the importer's name and address appear on the device, packaging or accompanying document. Distributors must verify marking and declarations before making a device available and must escalate complaints.
| Actor | Core obligation | MDR basis |
|---|---|---|
| Manufacturer, inside or outside the EU | Conformity assessment, technical file, declaration, CE marking, post-market surveillance, vigilance | Articles 10 and 83 to 86 |
| Authorised representative in the EU | Verify declaration and file availability, act as local contact, forward incidents | Article 11 |
| Importer | Verify marking and declaration, register, label, keep complaints | Article 13 |
| Distributor | Preserve storage and transport conditions, verify marking, escalate complaints | Article 14 |
Source: Regulation (EU) 2017/745, Articles 10 to 15
Two failure modes repeat in practice. The first is a shipment where the CE mark exists but the declaration names a different legal entity, usually because a trading company issued the document. The second is a declaration that covers the family but not the specific model with the added accessory. Both are document failures, resolved before dispatch rather than at the border.
After CE marking, the manufacturer also owes a post-market surveillance system and a vigilance process. Class I devices require a post-market surveillance report, while Class IIa devices and above require the periodic safety update report (EU MDR 2017/745, Articles 85 and 86). An importer who receives no post-market information from a supplier has no early warning path for field complaints.
FAQ
Is a rollator a Class I medical device under EU MDR?
Most rollators and walkers are Class I, non-invasive devices with no measuring function, assessed under Annex VIII Rule 1. A model with a measuring function becomes Class Im and needs notified body involvement for that function. The classification depends on the intended purpose the manufacturer states, so confirm the rule applied in the technical file rather than on the quotation.
Does a Class I walker need a notified body certificate?
No, when the device is non-sterile and has no measuring function. The manufacturer carries out the conformity assessment, compiles the technical documentation and signs the Declaration of Conformity without third-party review. Notified body involvement appears only for Class Is and Class Im configurations, or when a manufacturer voluntarily uses a body.
Is a CE certificate the same as a EU Declaration of Conformity?
No. A Declaration of Conformity is the manufacturer's signed legal statement that the device meets EU MDR 2017/745, and every device needs one. A certificate is issued by a notified body and exists only for classes that require third-party assessment. For most walkers, the declaration is the document to request.
How long must the technical documentation be kept?
The manufacturer must keep the technical documentation, the declaration and related records available for at least ten years after the last device covered by the declaration has been placed on the market (EU MDR 2017/745, Article 10). For an importer, the practical question is whether the current revision can be produced within days of a request.
Must a Chinese manufacturer appoint an EU authorised representative?
Yes. A manufacturer established outside the Union must appoint a single authorised representative in the Union before placing a device on the market (EU MDR 2017/745, Article 11). The mandate covers verification of the declaration and technical file, cooperation with authorities and incident handling. The representative's details must be available to the importer and on request to authorities.
What this guide does not cover
- This guide does not give a legal or compliance conclusion for any specific model, market or tender. Classification, documentation sufficiency and labelling compliance are judged by the manufacturer and by the buyer's regulatory advisor.
- UKCA marking, Swissmedic requirements and non-EU national registration routes are outside scope, even where the same device is sold in those markets.
- Sterile configurations, devices with a measuring function and software features are mentioned only to explain when notified body involvement appears; they are not assessed here.
If you are preparing a tender or a first shipment, send the model list and the target member states. We reply with the technical file index, the declaration status per model and the production lead time within one business day. You can also request a quote with your quantities and packaging requirements, or review the rollator walkers range and our quality and compliance documentation before deciding which models to shortlist.
