A Class I device declared under EU MDR without a notified body still appears in EUDAMED, and self-declaration removes the certificate step, not the data entry. The first four modules became mandatory on 28 May 2026, and devices already on the market must be registered by late November 2026 to stay listed. This guide lists fields, owners and gaps.
Which EUDAMED Modules Apply to a Class I Manufacturer?
Four of the six modules have been mandatory since 28 May 2026, and a Class I self-declaring manufacturer uses two of them routinely.
Actor registration comes first and produces a Single Registration Number. Without it the device module has nothing to attach the device record to, so this is the step that gates everything else.
UDI and device registration comes second. It carries the device record itself, linked to the Basic UDI-DI and to the manufacturer's registration, and it is the module a distributor will search when checking whether a product is legitimately on the EU market.
The certificates module matters mainly to manufacturers who hold a notified body certificate, which a self-declared Class I device does not. Its absence from a Class I file is expected rather than a gap.
The market surveillance module is used by authorities rather than by manufacturers, though a manufacturer may be asked to respond to information recorded there.
The two remaining modules, on vigilance and on clinical investigation, are not yet mandatory, and their status should be checked rather than assumed.
This is the filing half of the compliance file that sits beside the test report and the declaration of conformity, and the three are read together by a distributor checking a product. That does not remove the underlying obligations, and a manufacturer should not read the module timetable as a timetable for reporting.
| Module | Mandatory since | Class I use |
|---|---|---|
| Actors | 28 May 2026 | Register the manufacturer |
| UDI and devices | 28 May 2026 | Register each device |
| Certificates | 28 May 2026 | Not used by self-declaration |
| Market surveillance | 28 May 2026 | Authority-side |
Source: Regulation (EU) 2017/745; European Commission EUDAMED pages, 2026
What Data Do You Have to Enter?
The device record is a structured dataset, and most rejected submissions fail on fields the manufacturer considered administrative rather than technical.
The identity block carries the trade name, the basic model description and the EMDN nomenclature code. The code is chosen from a published list rather than written freely, and choosing a neighbouring code is the most common cause of a record that cannot be matched later.
The classification block carries the risk class and the status flags for sterile, measuring and reusable surgical instrument. For a Class I device all three flags are usually negative, and a wrong flag changes the conformity route rather than only the data.
The UDI block carries the Basic UDI-DI, which identifies the model family, and the UDI-DI for each packaging configuration. A product in three pack sizes has three UDI-DIs and one Basic UDI-DI, and treating them as one is a recurring error.
The economic operator block links the manufacturer, any authorised representative and, where relevant, the importer. The links are to registration numbers rather than to free text, which is why the actor step cannot be skipped.
The packaging block records the levels at which the UDI carrier appears. A label applied only to the outer carton leaves the unit-level carrier missing, and that shows up when a hospital scans a single device. Record which level carries the carrier for every configuration, because the answer changes with pack design.
| Group | What goes in | Held by |
|---|---|---|
| Identity | Trade name, EMDN | Regulatory |
| Classification | Class and flags | Regulatory |
| UDI | Basic UDI-DI and DI | Labelling |
| Operators | Manufacturer and AR | Regulatory |
| Packaging | Carrier levels | Packaging |
Source: EUDAMED data model; internal data ownership review, 2026 (company statement)
Who Owns Which Field?
The data entry is rarely one person's job, and the failures cluster where ownership was assumed rather than assigned.
Regulatory affairs owns the identity, classification and operator blocks, because those fields follow the technical documentation and the declaration of conformity. Whoever signs the declaration should be able to explain every value in these blocks.
Labelling and packaging own the UDI block and the carrier levels, because those fields are physical decisions made on the artwork. When labelling changes a carton, the device record has to change with it, and nothing in the system forces that conversation.
Quality owns the link between the record and the technical file. A device record that names a model the technical file does not cover is a finding waiting to happen, and it is usually caused by a product manager adding a variant.
Commercial teams create the most avoidable gaps, because a new pack size or a bundled kit is a sales decision before it is a regulatory one. A rule that no new configuration ships without a UDI-DI is cheaper than a correction later.
A single named owner for the device record, backed by those four contributors, is the arrangement that survives staff changes, and it is worth naming a deputy for the same reason. The register of changes that trigger an update should sit with that owner rather than in a project folder. Ownership by committee produces a record that is complete only while the committee still meets.
What Goes Wrong Most Often?
The recurring failures are administrative, they are cheap to prevent, and each one delays a registration that a manufacturer thought was finished.
The first is a mismatch between the Basic UDI-DI and the model family in the technical file. The database accepts the record, the documentation does not match it, and the discrepancy surfaces at a market surveillance enquiry rather than at submission.
The second is a missing authorised representative link. A non-EU manufacturer without a valid mandate cannot complete the operator block, and the record sits incomplete while the product is already being sold.
The third is stale data after a change. A new supplier of a component, a new pack size or a renamed model all require an update, and the record is usually updated only when someone remembers.
The fourth is the legacy deadline itself, which is the one most likely to catch a distributor rather than a manufacturer. Devices placed on the market before 28 May 2026 must be registered to stay listed, with a deadline at the end of November 2026 and 28 November cited in most guidance. Sources differ by a day, so confirm the date on the European Commission page rather than in a supplier's summary.
| Gap | Consequence | Fix |
|---|---|---|
| Family mismatch | Surveillance finding | Reconcile to file |
| Missing AR | Incomplete record | Valid mandate |
| Stale data | Withdrawn listing | Change trigger |
| Legacy timing | Late registration | Register now |
Source: EUDAMED data model; internal registration review, 2026 (company statement)
Frequently Asked Questions
Does a Class I self-declared device need a notified body for EUDAMED?
No. Self-declaration removes the notified body from the conformity route, so no certificate is issued and the certificates module stays empty. The actor and device modules still apply, and they are the two a Class I manufacturer will use. Budget the data entry as a project rather than an afternoon.
Is UDI the same as a barcode?
No. A barcode is a carrier, while the UDI is the identifier the carrier encodes, made of a device identifier and a production identifier. The same UDI-DI can appear as a linear barcode or as a data matrix, and the choice belongs to the labelling decision. Printing method follows the surface, not the identifier.
What is the difference between Basic UDI-DI and UDI-DI?
The Basic UDI-DI identifies a model family and is used in the technical documentation and the declaration; the UDI-DI identifies a specific product and packaging configuration and is what appears on the label. One family can carry several UDI-DIs, which is why a change of pack size creates a record change.
Who enters the data when the manufacturer is outside the EU?
The manufacturer remains responsible for its own record and usually enters it directly. The authorised representative is linked to that record rather than replacing it, and an importer appears as a separate operator with its own registration. Confirm the mandate in writing before the record is submitted.
What happens if the record is not updated after a change?
The listing describes a product that no longer exists in that form. The practical risk is a mismatch discovered during a surveillance enquiry or a customs check, and the remedy is a correction that could have been an edit. Assign one owner to watch for change triggers.
What This Guide Does Not Cover
This guide does not register any device, does not interpret the regulation for a specific product, and does not replace the official EUDAMED guidance or legal advice. Module statuses and deadlines have changed more than once and should be confirmed on the European Commission pages before a submission is planned. Classification decisions and data-model details belong to the manufacturer's regulatory function.
For products in the Class I range, start with the walking canes in the Class I range and their technical files. How documentation is kept current is described in the quality and compliance process. Send your device list and target market to request a quote; we reply within one business day.